FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 4163538
·
Received October 10, 2014
Report
- Report Number
- 2032227-2014-36410
- Event Type
- Malfunction
- Date Received
- October 10, 2014
- Date of Event
- September 11, 2014
- Report Date
- September 11, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
Additional Manufacturer Narrative · 1
A CRACKED RESERVOIR TUBE LIP AND MINOR SCRATCHES ON THE DISPLAY WINDOW WERE NOTED DURING THE VISUAL INSPECTION. NO BUTTON ERROR ALARMS WERE NOTED. ALL OF THE BUTTONS FUNCTIONED PROPERLY. HOWEVER, MOISTURE DAMAGE WAS NOTED ON THE KEYPAD TRACES.
Description of Event or Problem · 1
CUSTOMER REPORTED THAT THE INSULIN PUMP RETURNED A BUTTON ERROR ALARM. CUSTOMER DID NOT RECALL ANY SIGNIFICANT EVENTS LEADING UP TO THE ERROR ALARM. BLOOD GLUCOSE LEVEL WAS 144 MG/DL AT THE TIME OF THE CALL. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 639116 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-723LNAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 64 YR |