FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4163508 · Received October 10, 2014

Report

Report Number
2032227-2014-36399
Event Type
Malfunction
Date Received
October 10, 2014
Date of Event
September 11, 2014
Report Date
September 11, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM DURING TESTING. THE DEVICE HAD CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, MINOR SCRATCHED LCD WINDOW, SCRATCHED RESERVOIR TUBE WINDOW, BROKEN BELT CLIP SLOT, AND STAINED END CAP STICKER.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4)

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THE INSULIN PUMP RETURNED A BUTTON ERROR ALARM. CUSTOMER REPORTED THAT THE BUTTONS ON THE INSULIN PUMP'S KEYPAD WERE UNRESPONSIVE. CUSTOMER STATED THAT THE INSULIN PUMP MAY HAVE RECENTLY BEEN EXPOSED TO SWEAT. BLOOD GLUCOSE LEVEL WAS 78 MG/DL AT THE TIME OF THE CALL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
639735 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-722LNAB

Patients

Seq Age Sex Outcome Treatment
1 37 YR