FDA Adverse Event
Malfunction
Summary report: N
I-STAT1 ANALYZER, IMMUNO READY
MDR report key: 4163492
·
Received October 10, 2014
Report
- Report Number
- 2245578-2014-00067
- Event Type
- Malfunction
- Date Received
- October 10, 2014
- Date of Event
- July 1, 2014
- Report Date
- October 8, 2014
- Manufacturer
- ABBOTT POINT OF CARE
- Product Code
- CGA
- PMA / PMN Number
- K001387
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
(B)(4).
Description of Event or Problem · 1
ON (B)(4) 2014, ABBOTT POINT OF CARE (APOC) DISTRIBUTOR REPAIR CENTER WAS CONTACTED BY A CUSTOMER WHO REPORTED THAT ANALYZER SN (B)(4) WOULD NOT ACTIVATE. THE ANALYZER WAS RETURNED FOR REPAIR AND DURING FAILURE ANALYSIS PREFORMED ON (B)(4) 2014 A BURNT CAPACITOR (C26) WAS DISCOVERED. APOC HAS DETERMINED THAT THE COMPONENT FAILURE IDENTIFIED WITHIN THE ANALYZER CIRCUITRY MAY LEAD TO THE BATTERIES BECOMING UNCOMFORTABLY HOT TO TOUCH IN THE AREA OF THE BATTERY COMPARTMENT WHEN USING A GREEN NON-FUSED BATTERY CARRIER. BASED ON THE INFORMATION AVAILABLE THERE WERE NO PATIENT OR USER RELATED INJURIES ASSOCIATED WITH THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 639092 | I-STAT1 ANALYZER, IMMUNO READY | I-STAT1 ANALYZER | CGA | ABBOTT POINT OF CARE | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |