FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS

MDR report key: 4163460 · Received October 10, 2014

Report

Report Number
2649622-2014-12324
Event Type
Malfunction
Date Received
October 10, 2014
Date of Event
July 24, 2014
Report Date
July 24, 2014
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE DISTAL LV (LOW VOLTAGE) ELECTRODE OF THE LEAD WAS COVERED IN BLOOD AND BODY TISSUE/FIBROTIC GROWTH, THE HELIX OF THE LEAD BECAME EXTRINSICALLY DISTORTED DUE TO PULLING/STRETCHING/OVERSTRESS, AND VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING IMPLANT THE HELIX OF THE RIGHT ATRIAL (RA) LEAD WOULD NOT RETRACT FULLY. THE LEAD WAS NOT USED AND ANOTHER LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
639511 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5076-45

Patients

Seq Age Sex Outcome Treatment
1 00059 YR