530G INSULIN PUMP
Report
- Report Number
- 2032227-2014-36071
- Event Type
- Malfunction
- Date Received
- October 10, 2014
- Date of Event
- September 4, 2014
- Report Date
- September 11, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZO
- PMA / PMN Number
- 120010
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY
- Reporter Occupation
- PATIENT
Narratives
A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING, IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).
IT WAS REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR. CUSTOMER'S BLOOD GLUCOSE READING WAS 120 MG/DL. NO SIGNIFICANT EVENTS LEADING TO THE ALARM WERE OBSERVED. IT WAS REPORTED THAT CUSTOMER WAS UNABLE TO REWIND THE INSULIN PUMP. ADVISED DISCONTINUATION OF THE INSULIN PUMP. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 639525 | 530G INSULIN PUMP | OZO | OZO | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-751NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 52 YR |