FDA Adverse Event
Malfunction
Summary report: N
ATTAIN ABILITY PLUS
MDR report key: 4163420
·
Received October 10, 2014
Report
- Report Number
- 2649622-2014-12330
- Event Type
- Malfunction
- Date Received
- October 10, 2014
- Date of Event
- July 24, 2014
- Report Date
- July 24, 2014
- Manufacturer
- MPRI
- Product Code
- OJX
- PMA / PMN Number
- P080006
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- RI, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED. ANALYSIS WAS PERFORMED AND NO ANOMALIES WERE FOUND. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING AN IMPLANT PROCEDURE, THE LEFT VENTRICULAR (LV) LEAD BECAME DISLODGED AND COULD NOT BE SECURED TO THE PATIENT'S ANATOMY. THE PROCEDURE TIME WAS LENGTHENED BY APPROXIMATELY TWO HOURS. THE LV LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 640264 | ATTAIN ABILITY PLUS | DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE | OJX | MPRI | 429678 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00057 YR |