FDA Adverse Event Malfunction Summary report: N

PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE

MDR report key: 4163415 · Received October 10, 2014

Report

Report Number
2183613-2014-01244
Event Type
Malfunction
Date Received
October 10, 2014
Date of Event
May 12, 2014
Report Date
July 25, 2014
Manufacturer
MEDTRONIC MILACA, INC.
Product Code
LWP
PMA / PMN Number
P820003
Removal / Correction Number
Z-1661-2014
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: ANALYSIS DETERMINED THAT THE UPPER AND LOWER CASES WERE BROKEN AND THAT THE MAIN PRINTED CIRCUIT BOARD WAS OUT OF SPECIFICATION ELECTRICALLY AND THAT IT NEEDED INTERCONNECT WASHERS. (B)(4).

Additional Manufacturer Narrative · 1

THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION, BUT RETURNED PRODUCT TESTING FOUND THE DEVICE DID NOT PERFORM AS DESCRIBED IN THE FIELD ACTION. FURTHER ANALYSIS WAS PERFORMED ON THE MAIN PRINTED CIRCUIT BOARD (PCB). VISUAL INSPECTION REVEALED NO ANOMALIES. BENCH ANALYSIS CONFIRMED THE CURRENT DRAIN FUNCTIONAL TEST ISSUE. HOWEVER ANALYSIS WAS HALTED DUE TO COMPLEXITY OF POSSIBLE CAUSE. CONCLUSION: CONFIRMED THE ACTIVE CURRENT DRAIN FAILURE, BUT COULD NOT SUCCESSFULLY ISOLATE THE FAULT DUE TO FAILURE COMPLEXITY.

Description of Event or Problem · 1

THE EXTERNAL PULSE GENERATOR WAS RETURNED AS A TRADE-IN/EXCHANGE AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
639827 PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE LWP MEDTRONIC MILACA, INC. 5348

Patients

Seq Age Sex Outcome Treatment
1