CAPSUREFIX NOVUS
Report
- Report Number
- 2649622-2014-12379
- Event Type
- Injury
- Date Received
- October 10, 2014
- Date of Event
- August 16, 2014
- Report Date
- August 20, 2014
- Manufacturer
- MPRI
- Product Code
- DTB
- PMA / PMN Number
- P930039
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
PRODUCT EVENT SUMMARY : THE DEVICE WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY INDICATED OVERSENSING DUE TO NON-PHYSIOLOGIC SIGNALS/SIC (SENSING INTEGRITY COUNTER).
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE EMERGENCY ROOM DUE TO A LEAD INTEGRITY ALERT. THE RIGHT VENTRICULAR (RV) LEAD DEMONSTRATED OVERSENSING. THE RIGHT ATRIAL (RA) LEAD SHOWED HIGH IMPEDANCE AND OVERSENSING DUE TO POSSIBLE FRACTURE. BOTH LEADS WERE CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 644009 | CAPSUREFIX NOVUS | ELECTRODE, PACEMAKER, PERMANENT | DTB | MPRI | 5076-52 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00064 YR | Hospitalization| R | 419688 LEAD, 694765 LEAD, D224TRK ICD |