CAPSUREFIX NOVUS
Report
- Report Number
- 2649622-2014-12434
- Event Type
- Injury
- Date Received
- October 10, 2014
- Date of Event
- June 10, 2014
- Report Date
- August 19, 2014
- Manufacturer
- MPRI
- Product Code
- DTB
- PMA / PMN Number
- P930039
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
THIS REPORT IS BASED SOLELY ON DEVICE ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED IN SEGMENTS. THE SEGMENTS WERE ANALYZED AND IT WAS FOUND THAT THE DISTAL CONDUCTOR OF THE LEAD DEVELOPED A FRACTURE DUE TO FLEXING WHILE IN VIVO. CONCOMITANT MEDICAL PRODUCTS: MODEL: 7000; COMPETITOR LEAD; IMPLANT: (B)(6) 2008. (B)(4).
IT WAS REPORTED THAT THE LEAD WAS EXPLANTED DUE TO A SYSTEM UPGRADE. THE LEAD WAS RETURNED TO THE MANUFACTURER, ANALYZED, AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 644355 | CAPSUREFIX NOVUS | ELECTRODE, PACEMAKER, PERMANENT | DTB | MPRI | 5076-58 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00082 YR | Hospitalization| R | D154VWC, ICD |