FDA Adverse Event Malfunction Summary report: N

PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE

MDR report key: 4163160 · Received October 10, 2014

Report

Report Number
2182208-2014-03007
Event Type
Malfunction
Date Received
October 10, 2014
Date of Event
August 7, 2014
Report Date
August 22, 2014
Manufacturer
MEDTRONIC, INC.
Product Code
LWP
PMA / PMN Number
P820003
Removal / Correction Number
Z-1661-2014
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION, RETURNED PRODUCT TESTING FOUND THE DEVICE DID PERFORM AS DESCRIBED IN THE FIELD ACTION. PRODUCT EVENT SUMMARY: ANALYSIS FOUND THE DEVICE FAILED INCOMING FUNCTIONAL TESTING, FUNCTIONAL TEST FAILURES RERAN TEN TIMES AND THE DEVICE PASSED (INTERMITTENT OPERATION). ANALYSIS ALSO FOUND UPPER CASE, LOWER CASE, AND ONE SIDE BAIL COVER ARE BROKEN AND CONTAMINATED. (B)(4).

Description of Event or Problem · 1

THE EXTERNAL PULSE GENERATOR WAS RETURNED TO THE MANUFACTURER, ANALYZED AND TESTED OUT OF SPECIFICATION. NO PATIENT INVOLVEMENT OR COMPLICATIONS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
644887 PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE LWP MEDTRONIC, INC. 5348