PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE
Report
- Report Number
- 2182208-2014-03007
- Event Type
- Malfunction
- Date Received
- October 10, 2014
- Date of Event
- August 7, 2014
- Report Date
- August 22, 2014
- Manufacturer
- MEDTRONIC, INC.
- Product Code
- LWP
- PMA / PMN Number
- P820003
- Removal / Correction Number
- Z-1661-2014
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
THIS REPORT IS BASED SOLELY ON DEVICE ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION, RETURNED PRODUCT TESTING FOUND THE DEVICE DID PERFORM AS DESCRIBED IN THE FIELD ACTION. PRODUCT EVENT SUMMARY: ANALYSIS FOUND THE DEVICE FAILED INCOMING FUNCTIONAL TESTING, FUNCTIONAL TEST FAILURES RERAN TEN TIMES AND THE DEVICE PASSED (INTERMITTENT OPERATION). ANALYSIS ALSO FOUND UPPER CASE, LOWER CASE, AND ONE SIDE BAIL COVER ARE BROKEN AND CONTAMINATED. (B)(4).
THE EXTERNAL PULSE GENERATOR WAS RETURNED TO THE MANUFACTURER, ANALYZED AND TESTED OUT OF SPECIFICATION. NO PATIENT INVOLVEMENT OR COMPLICATIONS WERE REPORTED.