SPRINT QUATTRO
Report
- Report Number
- 2649622-2014-12500
- Event Type
- Injury
- Date Received
- October 10, 2014
- Date of Event
- July 24, 2014
- Report Date
- July 24, 2014
- Manufacturer
- MPRI
- Product Code
- LWS
- PMA / PMN Number
- P920015
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT EVENT SUMMARY: THE LEAD WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY INDICATED THE IMPEDANCE ON THE RV (RIGHT VENTRICULAR) PACING LEAD WAS BEYOND THE EXPECTED UPPER RANGE, LEAD WARNING ALERT, AND OVERSENSING DUE TO NON-PHYSIOLOGIC SIGNALS/SIC (SENSING INTEGRITY COUNTER). AN RV PACE LEAD IMPEDANCE ALERT OCCURRED ON (B)(6) 2014. MAX VENTRICULAR PACE IMPEDANCE RISES FROM 736 OHM THE WEEK ENDING (B)(6) 2014 TO GREATER THAN 2300 OHM THE WEEK ENDING (B)(6) 2014. THERE ARE 4 NON-SUSTAINED TACHYCARDIA (NST) EVENTS OF LESS THAN 220 MS V-V CYCLE RECORDED ON (B)(6) 2014.
THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THE PATIENT WENT TO THE HOSPITAL DUE TO AN AUDIBLE ALERT. IT WAS NOTED THERE WAS SHORT V-V INTERVALS, HIGH IMPEDANCE, SOME NON-SUSTAINED VENTRICULAR TACHYCARDIAS AND CONFIRMED FRACTURE. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 639068 | SPRINT QUATTRO | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | MPRI | 694465 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R |