SPRINT QUATTRO
Report
- Report Number
- 2649622-2014-12516
- Event Type
- Malfunction
- Date Received
- October 10, 2014
- Report Date
- July 25, 2014
- Manufacturer
- MPRI
- Product Code
- LWS
- PMA / PMN Number
- P920015
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT EVENT SUMMARY: THE LEAD WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY INDICATED THE IMPEDANCE ON THE RV (RIGHT VENTRICULAR) PACING LEAD WAS BEYOND THE EXPECTED UPPER RANGE AND THE IMPEDANCE TREND ON THE RV (RIGHT VENTRICULAR) PACING LEAD WAS RISING. MAX RV PACE IMPEDANCE RISES FROM 768 OHM THE WEEK ENDING (B)(4) 2013 TO GREATER THAN 1900 OHM THE WEEK ENDING (B)(4) 2014.
THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THERE WAS HIGH PACING IMPEDANCE AND POSSIBLE FRACTURE ON THE RIGHT VENTRICULAR (RV) LEAD. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.