FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO

MDR report key: 4163008 · Received October 10, 2014

Report

Report Number
2649622-2014-12520
Event Type
Injury
Date Received
October 10, 2014
Date of Event
July 30, 2014
Report Date
July 30, 2014
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE LEAD WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY INDICATED THE IMPEDANCE ON THE RV (RIGHT VENTRICULAR) PACING LEAD WAS BEYOND THE EXPECTED UPPER RANGE AND LEAD IMPEDANCE OUT OF RANGE ALERT. BEGINNING ON (B)(6) 2014, RV BIPOLAR LEAD IMPEDANCE MEASURED 2869 OHMS RISING TO OVER 3000 OHMS (B)(6) 2014. RV LEAD IMPEDANCE WARNING OCCURRED ON (B)(6) 2014 FOR IMPEDANCE GREATER THAN 3000 OHMS.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS HIGH IMPEDANCE AND POSSIBLE FRACTURE. THE LEAD WILL BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
639749 SPRINT QUATTRO DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 6944-65

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R