FDA Adverse Event
Injury
Summary report: N
CAPSUREFIX MRI SURESCAN
MDR report key: 4163003
·
Received October 10, 2014
Report
- Report Number
- 2649622-2014-12518
- Event Type
- Injury
- Date Received
- October 10, 2014
- Report Date
- July 28, 2014
- Manufacturer
- MPRI
- Product Code
- NVN
- PMA / PMN Number
- P090013
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED AND THE OUTER INSULATION OF THE LEAD DEVELOPED A BREACH DUE TO A DEPRESSION WHILE IN VIVO. THERE WAS BLOOD ON THE PROXIMAL CONDUCTOR OF THE LEAD AND IT WAS NOT OBSTRUCTED. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT THERE WAS HIGH AND RISING THRESHOLD. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 639376 | CAPSUREFIX MRI SURESCAN | DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE | NVN | MPRI | 5086MRI58 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00033 YR | Hospitalization| R |