FDA Adverse Event Injury Summary report: N

CAPSUREFIX MRI SURESCAN

MDR report key: 4163003 · Received October 10, 2014

Report

Report Number
2649622-2014-12518
Event Type
Injury
Date Received
October 10, 2014
Report Date
July 28, 2014
Manufacturer
MPRI
Product Code
NVN
PMA / PMN Number
P090013
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED AND THE OUTER INSULATION OF THE LEAD DEVELOPED A BREACH DUE TO A DEPRESSION WHILE IN VIVO. THERE WAS BLOOD ON THE PROXIMAL CONDUCTOR OF THE LEAD AND IT WAS NOT OBSTRUCTED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS HIGH AND RISING THRESHOLD. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
639376 CAPSUREFIX MRI SURESCAN DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE NVN MPRI 5086MRI58

Patients

Seq Age Sex Outcome Treatment
1 00033 YR Hospitalization| R