FDA Adverse Event
Injury
Summary report: N
CAPSURE SP
MDR report key: 4162992
·
Received October 10, 2014
Report
- Report Number
- 2649622-2014-12481
- Event Type
- Injury
- Date Received
- October 10, 2014
- Date of Event
- July 22, 2014
- Report Date
- July 22, 2014
- Manufacturer
- MPRI
- Product Code
- DTB
- PMA / PMN Number
- P830061
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: PRODUCT ID 4024 LEAD, IMPLANTED: (B)(6) 2000. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD LOW PACING IMPEDANCE. THE RV LEAD REMAINS IN USE. IT WAS FURTHER REPORTED THAT THE RIGHT ATRIAL (RA) LEAD HAD NO CAPTURE AT 8 VOLTS AND LOW PACING IMPEDANCE. THE RA LEAD WAS TURNED OFF AND REMAINS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 640193 | CAPSURE SP | ELECTRODE, PACEMAKER, PERMANENT | DTB | MPRI | 4524 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00058 YR | Hospitalization| R | VEDR01 IPG |