PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE
Report
- Report Number
- 2183613-2014-01259
- Event Type
- Malfunction
- Date Received
- October 10, 2014
- Date of Event
- May 22, 2014
- Report Date
- August 7, 2014
- Manufacturer
- MEDTRONIC MILACA, INC.
- Product Code
- LWP
- PMA / PMN Number
- P820003
- Removal / Correction Number
- Z-1661-2014
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
THIS REPORT IS BASED SOLELY ON DEVICE ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: ANALYSIS DETERMINED THAT THE UPPER CASE WAS BROKEN AND CONTAMINATED, THAT THE BATTERY DRAWER AND O-RING ARE CONTAMINATED AND THAT THE BATTERY RECEIVED WITH THE DEVICE WAS CORRODED. THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION, BUT RETURNED PRODUCT TESTING FOUND THE DEVICE DID NOT PERFORM AS DESCRIBED IN THE FIELD ACTION. (B)(4).
THE EXTERNAL PULSE GENERATOR WAS RETURNED AS A TRADE-IN/EXCHANGE AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 639621 | PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE | LWP | MEDTRONIC MILACA, INC. | 5348 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |