MARQUIS VR
Report
- Report Number
- 9614453-2014-02492
- Event Type
- Injury
- Date Received
- October 10, 2014
- Date of Event
- July 29, 2014
- Report Date
- July 29, 2014
- Manufacturer
- IPG MFG SWITZERLAND
- Product Code
- LWS
- PMA / PMN Number
- P980016
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- PHYSICIAN
Narratives
THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. THE DEVICE MET 91% OF EXPECTED LONGEVITY. WITHOUT THE HISTORY OF THE PROGRAMMED SETTINGS THROUGHOUT ITS SERVICE LIFE, THERE IS NO WAY TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL. (B)(4).
IT WAS REPORTED THAT DURING THE INTERROGATION OF THE DEVICE, THE DEVICE COULD NOT BE TELEMETERED. IT WAS NOTED THE DEVICE HAD REACHED BATTERY DEPLETION. THE PHYSICIAN REPORTED THE EVENT AS PREMATURE BATTERY DEPLETION. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 639260 | MARQUIS VR | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | IPG MFG SWITZERLAND | 7230CX |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R |