FDA Adverse Event Malfunction Summary report: N

CARELINK

MDR report key: 4162929 · Received October 10, 2014

Report

Report Number
2182208-2014-03042
Event Type
Malfunction
Date Received
October 10, 2014
Date of Event
August 27, 2014
Report Date
August 27, 2014
Manufacturer
MEDTRONIC, INC.
Product Code
KRG
PMA / PMN Number
P890003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 1

CORRECTED INFORMATION NO EVAL EXPLAIN CODE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: ANALYSIS CONFIRMED THE CUSTOMER COMMENT THAT THE PROGRAMMER WOULD NOT POWER UP AND THE POWER SUPPLY WAS THEREFORE REPLACED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PROGRAMMER WOULD NOT POWER UP. THE USER TRIED ANOTHER POWER CORD AND ANOTHER OUTLET WITH NO DIFFERENT RESULT. TECHNICAL SUPPORT (TS) STATED THAT THE ISSUE IS MOST LIKELY A BAD POWER SUPPLY AND TO RETURN IT FOR REPAIR. THE PROGRAMMER HAS BEEN RETURNED FOR REPAIR. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
639048 CARELINK PROGRAMMER, PACEMAKER KRG MEDTRONIC, INC. 2090

Patients

Seq Age Sex Outcome Treatment
1 2290 PACING SYSTEM ANALYZER