FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS

MDR report key: 4162926 · Received October 10, 2014

Report

Report Number
2649622-2014-12554
Event Type
Malfunction
Date Received
October 10, 2014
Date of Event
August 14, 2014
Report Date
August 14, 2014
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY- THE FULL LEAD WAS RETURNED. THE DISTAL LV (LOW VOLTAGE) ELECTRODE OF THE LEAD WAS COVERED IN BLOOD, ANALYSIS WAS PERFORMED AND NO ANOMALIES WERE FOUND. THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE THE RIGHT VENTRICULAR (RV) LEAD SHOWED HIGH IMPEDANCE. THE PHYSICIAN CHOSE TO IMPLANT A DIFFERENT LEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
639028 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5076-52

Patients

Seq Age Sex Outcome Treatment
1 00073 YR