FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 4162892 · Received October 10, 2014

Report

Report Number
2649622-2014-12551
Event Type
Injury
Date Received
October 10, 2014
Date of Event
August 6, 2014
Report Date
August 6, 2014
Manufacturer
MPRI
Product Code
NVY
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT DISPLAYED T WAVE OVERSENSING ON THE RIGHT VENTRICULAR (RV) LEAD A WEEK AFTER A DEVICE CHANGE OUT. THE OVERSENSING WAS NOTED ON A DEVICE INTERROGATION. THE PHYSICIAN ELECTED TO REOPEN THE POCKET AND INVESTIGATE FURTHER. THE DEVICE WAS NOTED TO HAVE BLOOD IN THE RV LEAD HEADER PORT. THE DEVICE WAS EXPLANTED AND REPLACED AND THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
639011 SPRINT QUATTRO SECURE DRUG-ELUTING PERMANENT DEFIBRILLATOR ELECTRODES NVY MPRI 6947M55

Patients

Seq Age Sex Outcome Treatment
1 00079 YR Hospitalization| R DDBB1D4 ICD, 4076-45 LEAD