FDA Adverse Event
Injury
Summary report: N
SPRINT QUATTRO SECURE
MDR report key: 4162892
·
Received October 10, 2014
Report
- Report Number
- 2649622-2014-12551
- Event Type
- Injury
- Date Received
- October 10, 2014
- Date of Event
- August 6, 2014
- Report Date
- August 6, 2014
- Manufacturer
- MPRI
- Product Code
- NVY
- PMA / PMN Number
- P920015
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PATIENT DISPLAYED T WAVE OVERSENSING ON THE RIGHT VENTRICULAR (RV) LEAD A WEEK AFTER A DEVICE CHANGE OUT. THE OVERSENSING WAS NOTED ON A DEVICE INTERROGATION. THE PHYSICIAN ELECTED TO REOPEN THE POCKET AND INVESTIGATE FURTHER. THE DEVICE WAS NOTED TO HAVE BLOOD IN THE RV LEAD HEADER PORT. THE DEVICE WAS EXPLANTED AND REPLACED AND THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 639011 | SPRINT QUATTRO SECURE | DRUG-ELUTING PERMANENT DEFIBRILLATOR ELECTRODES | NVY | MPRI | 6947M55 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00079 YR | Hospitalization| R | DDBB1D4 ICD, 4076-45 LEAD |