SPRINT QUATTRO SECURE
Report
- Report Number
- 2649622-2014-12630
- Event Type
- Injury
- Date Received
- October 10, 2014
- Date of Event
- July 15, 2014
- Report Date
- July 15, 2014
- Manufacturer
- MPRI
- Product Code
- LWS
- PMA / PMN Number
- P920015
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: A 5076-52 LEAD, IMPLANTED (B)(6) 2008. (B)(4).
PRODUCT EVENT SUMMARY : THE DEVICE WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY INDICATED THE IMPEDANCE ON THE RV (RIGHT VENTRICULAR) PACING LEAD WAS BEYOND THE EXPECTED UPPER RANGE. ANALYSIS OF THE DEVICE MEMORY ALSO INDICATED OVERSENSING DUE TO NON-PHYSIOLOGIC SIGNALS/SIC (SENSING INTEGRITY COUNTER). BEGINNING (B)(6) 2014, 101 NON-PHYSIOLOGIC SENSING INTEGRITY COUNTS (SIC) WERE RECORDED. BEGINNING THE WEEK OF (B)(6) 2014, RV PACING IMPEDANCE BEGAN INCREASING TO A HIGH OF 1096 OHMS.
IT WAS REPORTED THAT OVERSENSING, SHORT V-V INTERVALS, AND HIGH IMPEDANCE WERE OBSERVED ON THE RIGHT VENTRICULAR (RV) LEAD. A LEAD FRACTURE WAS SUSPECTED. THE RV LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 643377 | SPRINT QUATTRO SECURE | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | MPRI | 694765 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00074 YR | Hospitalization| R | D154AWG ICD |