FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 4162728 · Received October 10, 2014

Report

Report Number
2649622-2014-12630
Event Type
Injury
Date Received
October 10, 2014
Date of Event
July 15, 2014
Report Date
July 15, 2014
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: A 5076-52 LEAD, IMPLANTED (B)(6) 2008. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY : THE DEVICE WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY INDICATED THE IMPEDANCE ON THE RV (RIGHT VENTRICULAR) PACING LEAD WAS BEYOND THE EXPECTED UPPER RANGE. ANALYSIS OF THE DEVICE MEMORY ALSO INDICATED OVERSENSING DUE TO NON-PHYSIOLOGIC SIGNALS/SIC (SENSING INTEGRITY COUNTER). BEGINNING (B)(6) 2014, 101 NON-PHYSIOLOGIC SENSING INTEGRITY COUNTS (SIC) WERE RECORDED. BEGINNING THE WEEK OF (B)(6) 2014, RV PACING IMPEDANCE BEGAN INCREASING TO A HIGH OF 1096 OHMS.

Description of Event or Problem · 1

IT WAS REPORTED THAT OVERSENSING, SHORT V-V INTERVALS, AND HIGH IMPEDANCE WERE OBSERVED ON THE RIGHT VENTRICULAR (RV) LEAD. A LEAD FRACTURE WAS SUSPECTED. THE RV LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
643377 SPRINT QUATTRO SECURE DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 694765

Patients

Seq Age Sex Outcome Treatment
1 00074 YR Hospitalization| R D154AWG ICD