FDA Adverse Event Injury Summary report: N

VIRTUOSO DR

MDR report key: 4162681 · Received October 10, 2014

Report

Report Number
9614453-2014-02510
Event Type
Injury
Date Received
October 10, 2014
Date of Event
July 17, 2014
Report Date
August 14, 2014
Manufacturer
IPG MFG SWITZERLAND
Product Code
LWS
PMA / PMN Number
P980016
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS REVEALED THE RETURNED DEVICE DID NOT DETECT ANY PERFORMANCE ISSUES, BUT THE CALCULATED LONGEVITY RESULT OF 81% OF EXPECTED WAS LESS THAN THE PREDICTED VALUE BASED ON THE AVAILABLE PROGRAMMED PARAMETERS. IN THE ABSENCE OF THE COMPLETE PROGRAMMING HISTORY, IT IS NOT POSSIBLE TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL.

Additional Manufacturer Narrative · 1

THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES; HOWEVER, IT IS SAME/SIMILAR TO A DEVICE MARKETED IN THE U.S. THIS EVENT OCCURRED OUTSIDE THE US AND PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE BATTERY POSSIBLY DEPLETED PREMATURELY. THE DEVICE WAS EXPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
643920 VIRTUOSO DR DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS IPG MFG SWITZERLAND D164AWG

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R