FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE S

MDR report key: 4162643 · Received October 10, 2014

Report

Report Number
2649622-2014-12683
Event Type
Malfunction
Date Received
October 10, 2014
Date of Event
August 4, 2014
Report Date
August 4, 2014
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. THE DISTAL LV (LOW VOLTAGE) ELECTRODE OF THE LEAD WAS COVERED IN BLOOD.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN ATTEMPTED IMPLANT, ONCE PLACED IN THE DESIRED POSITION, THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED INADEQUATE THRESHOLD. THE PHYSICIAN ATTEMPTED TO TEST A NEW POSITION, HOWEVER THE HELIX OF THE LEAD WOULD NOT EXTEND. THE PHYSICIAN ATTEMPTED OVER 20 TURNS WITH THE SAME RESULT. THE PHYSICIAN ELECTED TO REMOVE AND REPLACED THE RV LEAD SUCCESSFULLY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
643841 SPRINT QUATTRO SECURE S DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 693565

Patients

Seq Age Sex Outcome Treatment
1 00077 YR