FDA Adverse Event Injury Summary report: N

INSYNC

MDR report key: 4162496 · Received October 10, 2014

Report

Report Number
9614453-2014-02518
Event Type
Injury
Date Received
October 10, 2014
Date of Event
August 4, 2014
Report Date
August 4, 2014
Manufacturer
IPG MFG SWITZERLAND
Product Code
LWP
PMA / PMN Number
P010015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. CONCOMITANT MEDICAL PRODUCTS: COMPETITOR LEAD, IMPLANTED: (B)(6) 2000. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE EXPERIENCED A POWER ON RESET (POR) WHEN THE PATIENT UNDERWENT A COMPUTERIZED TOMOGRAPHY (CT) SCAN. THE DEVICE WAS REPROGRAMMED TO OPERATE IN THE ORIGINAL SETTINGS AND REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
644703 INSYNC PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE LWP IPG MFG SWITZERLAND 8040

Patients

Seq Age Sex Outcome Treatment
1 00091 YR Required Intervention 4193 LEAD