FDA Adverse Event
Malfunction
Summary report: N
CAPSUREFIX NOVUS
MDR report key: 4162466
·
Received October 10, 2014
Report
- Report Number
- 3008973940-2014-00432
- Event Type
- Malfunction
- Date Received
- October 10, 2014
- Date of Event
- August 11, 2014
- Report Date
- August 11, 2014
- Manufacturer
- MEDTRONIC SINGAPORE OPERATIONS
- Product Code
- DTB
- PMA / PMN Number
- P930039
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED AND THE HELIX OF THE LEAD WAS EXTRINSICALLY BENT. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE, THERE WAS A SCREW PROBLEM. THE LEAD WAS ATTEMPTED AND NOT USED. ANOTHER LEAD WAS USED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 639954 | CAPSUREFIX NOVUS | ELECTRODE, PACEMAKER, PERMANENT | DTB | MEDTRONIC SINGAPORE OPERATIONS | 5076-58 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |