FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS

MDR report key: 4162460 · Received October 10, 2014

Report

Report Number
3008973940-2014-00433
Event Type
Malfunction
Date Received
October 10, 2014
Date of Event
August 22, 2014
Report Date
August 22, 2014
Manufacturer
MEDTRONIC SINGAPORE OPERATIONS
Product Code
DTB
PMA / PMN Number
P930039
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4)

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED AND THE DISTAL CONDUCTOR OF THE LEAD WAS EXTRINSICALLY OVER-ROTATED. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE ENTRANCE OF THE IMPLANTABLE PACING LEAD (IPL) WAS TOO NARROW FOR INTRODUCING A STYLET GUIDE WIRE. THE IPL COULD NOT BE IMPLANTED AND WAS EXCHANGED FOR ANOTHER LEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
639952 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MEDTRONIC SINGAPORE OPERATIONS 5076-58

Patients

Seq Age Sex Outcome Treatment
1