FDA Adverse Event Injury Summary report: N

PROTECTA CRT-D

MDR report key: 4162449 · Received October 10, 2014

Report

Report Number
9614453-2014-02520
Event Type
Injury
Date Received
October 10, 2014
Date of Event
June 18, 2014
Report Date
August 14, 2014
Manufacturer
IPG MFG SWITZERLAND
Product Code
NIK
PMA / PMN Number
P010031
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IS
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS REVEALED THE RETURNED DEVICE INDICATED NORMAL BATTERY DEPLETION.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES, HOWEVER, IT IS SIMILAR TO A DEVICE MARKETED IN THE U.S. THE EVENT IS BEING REPORTED DUE TO AN ALLEGED MALFUNCTION. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE INDICATED RECOMMENDED REPLACEMENT TIME (RRT) EARLIER THAN EXPECTED. THE DEVICE WAS EXPLANTED AND REPLACED SUCCESSFULLY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
639796 PROTECTA CRT-D DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO NIK IPG MFG SWITZERLAND D364TRG

Patients

Seq Age Sex Outcome Treatment
1 00083 YR Hospitalization| R