FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX

MDR report key: 4162448 · Received October 10, 2014

Report

Report Number
2649622-2014-12770
Event Type
Malfunction
Date Received
October 10, 2014
Date of Event
July 15, 2014
Report Date
July 15, 2014
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED AND THE DISTAL CONDUCTOR OF THE LEAD WAS EXTRINSICALLY OVER-ROTATED. THE DISTAL LV (LOW VOLTAGE) ELECTRODE OF THE LEAD WAS COVERED IN BLOOD AND VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE IMPLANT OF A RIGHT ATRIAL (RA) LEAD, THE HELIX WOULD NOT DEPLOY. THE RA LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
639702 CAPSUREFIX ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5568-53

Patients

Seq Age Sex Outcome Treatment
1 00046 YR