CD HORIZON SPINAL SYSTEM
MDR report key: 4162119
·
Received October 10, 2014
Report
- Report Number
- 1030489-2014-03994
- Event Type
- Malfunction
- Date Received
- October 10, 2014
- Date of Event
- August 27, 2014
- Report Date
- September 11, 2014
- Manufacturer
- WARSAW ORTHOPEDICS
- Product Code
- KWP
- PMA / PMN Number
- K981676
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4): NEITHER THE DEVICE NOR APPLICABLE IMAGING FILMS WERE RETURNED TO THE MANUFACTURER FOR EVALUATION, THEREFORE, THE CAUSE OF THE EVENT CANNOT BE DETERMINED.
Additional Manufacturer Narrative · 1
ADDITIONAL INFO: MACROSCOPIC AND OPTICAL EXAMINATION REVEALED THREAD CREST AND FLANK DAMAGE; THIS DAMAGE APPEARS TO HAVE INITIATED AT THE START OF THE THREAD, AND IS CONSISTENT AROUND THE DAMAGED PORTION OF THE THREAD. THE ABOVE OBSERVATIONS ARE CONSISTENT WITH MISALIGNMENT OF THE MAS AND SET SCREW THREADS DURING CONSTRUCT ASSEMBLY.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PATIENT UNDERWENT A POSTERIOR SPINAL SURGICAL PROCEDURE. IT WAS REPORTED THAT THE SET SCREW SLIPPED DURING INSERTION INTO THE BONE SCREW. THE SCREW WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 644941 | CD HORIZON SPINAL SYSTEM | APPLIANCE, FIXATION, SPINAL INTERLAMINAL | KWP | WARSAW ORTHOPEDICS | NA | 0274719W |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00077 YR | BONE SCREW |