FDA Adverse Event Malfunction Summary report: N

SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

MDR report key: 4162076 · Received October 10, 2014

Report

Report Number
1416980-2014-35264
Event Type
Malfunction
Date Received
October 10, 2014
Date of Event
September 13, 2014
Report Date
September 15, 2014
Manufacturer
BAXTER HEALTHCARE CORPORATION
Product Code
FKX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TU
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AS THE CASSETTE WAS NOT RETURNED AND THE LOT NUMBER IS UNKNOWN, A DEVICE ANALYSIS CANNOT BE COMPLETED. HOWEVER, AN INCOMPLETE PRIME WAS REPORTED AND IS A KNOWN CAUSE OF THIS ALARM. ¿THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE,¿ WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE, PROVIDES STEP-BY-STEP INSTRUCTIONS FOR PROPERLY PRIMING THE DISPOSABLE SET AND SAYS TO VERIFY THAT THE PATIENT LINE IS PROPERLY PRIMED. IT WARNS THE USER NOT TO CONNECT TO THE PATIENT LINE UNLESS THE FLUID LEVEL IS AT OR NEAR THE CONNECTOR AT THE END OF THE DISPOSABLE SET PATIENT LINE. A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

THIS IS A REPORT OF A PATIENT WHO DID NOT FOLLOW THERAPY STEPS DURING PERITONEAL DIALYSIS (PD) THERAPY. THE PATIENT WAS CONNECTED TO THE HOMECHOICE PATIENT LINE BEFORE PRIMING WAS COMPLETE AND PRIOR TO THE ¿CONNECT YOURSELF¿ PROMPT. THE PATIENT WOULD CONTINUE THERAPY WITH ALL NEW SUPPLIES. THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE EVENT ASSOCIATED WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
643340 SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 HOMECHOICE