SD/PD LONG CURVED
Report
- Report Number
- 0001811755-2014-03552
- Event Type
- Malfunction
- Date Received
- October 10, 2014
- Date of Event
- September 11, 2014
- Report Date
- September 12, 2014
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- ERL
- PMA / PMN Number
- K112593
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OK, US
- Reporter Occupation
- OTHER
Narratives
THE REPORTED EVENT OF A BROKEN BUR WAS NOT CONFIRMED. DURING THE EVALUATION OF THE ORIGINALLY LISTED CONCOMITANT DEVICE ((B)(4) LOT 14099 - SD/PD LONG CURVED ATTACHMENT), THIS DEVICE COULD NOT ACCEPT A BUR DURING TESTING. IT WAS FOUND THAT THIS DEVICE HAD A DENTED BEARING GUIDE, A DENTED BEARING GUIDE WILL PREVENT A BUR FROM BEING INSERTED INTO THE ATTACHMENT. THE ASSOCIATED CUTTING ACCESSORY (UNKNOWN INS BUR, UNKNOWN LOT #) IS NOW LISTED AS THE CONCOMITANT DEVICE.
THE DEVICE IS NOT AVAILABLE FOR RETURN. A FOLLOW UP REPORT WILL BE FILED ONCE THE QUALITY INVESTIGATION IS COMPLETE. DEVICE NOT RETURNED TO MANUFACTURER.
IT WAS REPORTED THAT THE DRILL BIT BROKE OFF INSIDE OF THE ATTACHMENT DURING TESTING. IT WAS FURTHER REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT IN THIS EVENT.
IT WAS REPORTED THAT THE DRILL BIT BROKE OFF INSIDE OF THE ATTACHMENT DURING TESTING. IT WAS FURTHER REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT IN THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 643875 | SD/PD LONG CURVED | DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE | ERL | STRYKER INSTRUMENTS-KALAMAZOO | 14099 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN INS BUR, UNKNOWN INS LOT #| (B)(4)LOT 14099 |