FDA Adverse Event Malfunction Summary report: N

SD/PD LONG CURVED

MDR report key: 4162035 · Received October 10, 2014

Report

Report Number
0001811755-2014-03552
Event Type
Malfunction
Date Received
October 10, 2014
Date of Event
September 11, 2014
Report Date
September 12, 2014
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
ERL
PMA / PMN Number
K112593
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED EVENT OF A BROKEN BUR WAS NOT CONFIRMED. DURING THE EVALUATION OF THE ORIGINALLY LISTED CONCOMITANT DEVICE ((B)(4) LOT 14099 - SD/PD LONG CURVED ATTACHMENT), THIS DEVICE COULD NOT ACCEPT A BUR DURING TESTING. IT WAS FOUND THAT THIS DEVICE HAD A DENTED BEARING GUIDE, A DENTED BEARING GUIDE WILL PREVENT A BUR FROM BEING INSERTED INTO THE ATTACHMENT. THE ASSOCIATED CUTTING ACCESSORY (UNKNOWN INS BUR, UNKNOWN LOT #) IS NOW LISTED AS THE CONCOMITANT DEVICE.

Additional Manufacturer Narrative · 1

THE DEVICE IS NOT AVAILABLE FOR RETURN. A FOLLOW UP REPORT WILL BE FILED ONCE THE QUALITY INVESTIGATION IS COMPLETE. DEVICE NOT RETURNED TO MANUFACTURER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DRILL BIT BROKE OFF INSIDE OF THE ATTACHMENT DURING TESTING. IT WAS FURTHER REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT IN THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DRILL BIT BROKE OFF INSIDE OF THE ATTACHMENT DURING TESTING. IT WAS FURTHER REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT IN THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
643875 SD/PD LONG CURVED DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE ERL STRYKER INSTRUMENTS-KALAMAZOO 14099

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN INS BUR, UNKNOWN INS LOT #| (B)(4)LOT 14099