FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4161981 · Received October 10, 2014

Report

Report Number
2032227-2014-36022
Event Type
Injury
Date Received
October 10, 2014
Date of Event
September 10, 2014
Report Date
September 11, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE. (B)(4).

Description of Event or Problem · 1

CUSTOMER REPORTS TO HAVE A BLANK DISPLAY SCREEN. CUSTOMER'S BLOOD GLUCOSE WAS ABOUT 300 MG/DL, BUT STATES IT WAS 400 MG/DL THE PREVIOUS DAY. AFTER TROUBLESHOOTING, IT WAS FOUND THAT CUSTOMER HAD INCORRECT BATTERIES IN INSULIN PUMP. ONCE CUSTOMER CHANGED OUT BATTERIES, INSULIN PUMP DISPLAY SCREEN RETURNED. NO FURTHER INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
643788 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-523NAS

Patients

Seq Age Sex Outcome Treatment
1 60 YR Required Intervention