PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 4161981
·
Received October 10, 2014
Report
- Report Number
- 2032227-2014-36022
- Event Type
- Injury
- Date Received
- October 10, 2014
- Date of Event
- September 10, 2014
- Report Date
- September 11, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE. (B)(4).
Description of Event or Problem · 1
CUSTOMER REPORTS TO HAVE A BLANK DISPLAY SCREEN. CUSTOMER'S BLOOD GLUCOSE WAS ABOUT 300 MG/DL, BUT STATES IT WAS 400 MG/DL THE PREVIOUS DAY. AFTER TROUBLESHOOTING, IT WAS FOUND THAT CUSTOMER HAD INCORRECT BATTERIES IN INSULIN PUMP. ONCE CUSTOMER CHANGED OUT BATTERIES, INSULIN PUMP DISPLAY SCREEN RETURNED. NO FURTHER INFORMATION PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 643788 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-523NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR | Required Intervention |