SUPERDIMENSION INREACH SYSTEM
Report
- Report Number
- 3004962788-2014-00051
- Event Type
- Malfunction
- Date Received
- October 10, 2014
- Date of Event
- September 5, 2014
- Report Date
- May 27, 2015
- Manufacturer
- SUPERDIMENSION INC.
- Product Code
- JAK
- PMA / PMN Number
- K092365
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). DATE OF INITIAL REPORT : 10/10/2014. DATE OF FOLLOW-UP REPORT : 05/27/2015. AN ON-SITE SERVICE VISIT WAS PERFORMED AND A SYSTEM COMPONENT, THE LOCATION SUB-SYSTEM, WAS REPLACED DUE TO A BLOWN FUSE CAUSING AN OPEN CIRCUIT.
THE DEVICE WAS NOT RETURNED FOR EVALUATION. OUT OF AN ABUNDANCE OF CAUTION, SUPERDIMENSION IS FILING THIS MDR DUE TO THE ADDITIONAL RISK ASSOCIATED WITH MULTIPLE EXPOSURES TO GENERAL ANESTHESIA. THERE WERE NO ANOMALIES IDENTIFIED DURING THE INTERNAL REVIEW OF THE DHR OF THE SYSTEM CONSOLE. IF ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP REPORT WILL BE SUBMITTED.
SITE REPORTED THE SUPERDIMENSION ILOGIC SYSTEM DID NOT REGISTER THE PATIENT SENSOR TRIPLET AND THE PHYSICIAN CANCELLED THE SUPERDIMENSION PORTION OF THE CASE. THE PATIENT WAS UNDER GENERAL ANESTHESIA. THERE WAS NO REPORT OF PATIENT INJURY, DEATH OR OTHER SERIOUS ADVERSE EVENT. IF ADDITIONAL INFORMATION PERTINENT TO THE INCIDENT IS OBTAINED, A FOLLOW-UP REPORT WILL BE SUBMITTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 643702 | SUPERDIMENSION INREACH SYSTEM | ELECTROMAGNETIC NAVIGATION BRONCHOSCOPY | JAK | SUPERDIMENSION INC. | AAS00161-12 |
Patients