FDA Adverse Event Malfunction Summary report: N

SUPERDIMENSION INREACH SYSTEM

MDR report key: 4161961 · Received October 10, 2014

Report

Report Number
3004962788-2014-00051
Event Type
Malfunction
Date Received
October 10, 2014
Date of Event
September 5, 2014
Report Date
May 27, 2015
Manufacturer
SUPERDIMENSION INC.
Product Code
JAK
PMA / PMN Number
K092365
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DATE OF INITIAL REPORT : 10/10/2014. DATE OF FOLLOW-UP REPORT : 05/27/2015. AN ON-SITE SERVICE VISIT WAS PERFORMED AND A SYSTEM COMPONENT, THE LOCATION SUB-SYSTEM, WAS REPLACED DUE TO A BLOWN FUSE CAUSING AN OPEN CIRCUIT.

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR EVALUATION. OUT OF AN ABUNDANCE OF CAUTION, SUPERDIMENSION IS FILING THIS MDR DUE TO THE ADDITIONAL RISK ASSOCIATED WITH MULTIPLE EXPOSURES TO GENERAL ANESTHESIA. THERE WERE NO ANOMALIES IDENTIFIED DURING THE INTERNAL REVIEW OF THE DHR OF THE SYSTEM CONSOLE. IF ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

SITE REPORTED THE SUPERDIMENSION ILOGIC SYSTEM DID NOT REGISTER THE PATIENT SENSOR TRIPLET AND THE PHYSICIAN CANCELLED THE SUPERDIMENSION PORTION OF THE CASE. THE PATIENT WAS UNDER GENERAL ANESTHESIA. THERE WAS NO REPORT OF PATIENT INJURY, DEATH OR OTHER SERIOUS ADVERSE EVENT. IF ADDITIONAL INFORMATION PERTINENT TO THE INCIDENT IS OBTAINED, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
643702 SUPERDIMENSION INREACH SYSTEM ELECTROMAGNETIC NAVIGATION BRONCHOSCOPY JAK SUPERDIMENSION INC. AAS00161-12

Patients

Seq Age Sex Outcome Treatment
1