SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Report
- Report Number
- 1416980-2014-35284
- Event Type
- Malfunction
- Date Received
- October 10, 2014
- Date of Event
- September 16, 2014
- Report Date
- September 16, 2014
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). AS THE SAMPLE WAS NOT RETURNED AND THE LOT NUMBER IS UNKNOWN, A DEVICE ANALYSIS CANNOT BE COMPLETED. HOWEVER, A USE ERROR OF CONNECTING PRIOR TO THE COMPLETION OF PRIMING WAS REPORTED AND IS AN KNOWN CAUSE OF THE AIR. USE ERRORS AND PROPER USER INSTRUCTIONS ARE ADDRESSED IN ¿THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE¿, WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE. THE GUIDE PROVIDES STEP-BY-STEP INSTRUCTIONS FOR PROPERLY PRIMING THE DISPOSABLE SET. THE GUIDE WARNS THE USER NOT TO CONNECT TO YOUR PATIENT LINE UNLESS THE FLUID LEVEL IS AT OR NEAR THE CONNECTOR AT THE END OF THE DISPOSABLE SET PATIENT LINE. THE GUIDE INSTRUCTS THE USER TO VERIFY THAT THE PATIENT LINE IS PROPERLY PRIMED. THE GUIDE PROVIDES STEP-BY-STEP INSTRUCTIONS FOR PROPERLY REPRIMING THE PATIENT LINE. A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT THERE WAS AIR IN THE PATIENT LINE, DURING THE INITIAL DRAIN OF PERITONEAL DIALYSIS THERAPY; THE HOME PATIENT (HP) WAS CONNECTED. DURING TROUBLESHOOTING IT WAS DETERMINED THE HP HAD CONNECTED TO THE PATIENT LINE PRIOR TO THE COMPLETION OF PROPER PRIMING. THE TECHNICAL SERVICE REPRESENTATIVE ASSISTED THE HP IN ENDING THE THERAPY AND REVIEWED PROPER PROCEDURES WITH THE HP. THE HP WOULD COMPLETE THERAPY USING ALL NEW SUPPLIES. THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE EVENT ASSOCIATED WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 643665 | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 YR | HOMECHOICE |