FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4161921 · Received October 10, 2014

Report

Report Number
2032227-2014-36110
Event Type
Injury
Date Received
October 10, 2014
Date of Event
September 10, 2014
Report Date
September 11, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED THE DISPLACEMENT TEST, REWIND, BASIC OCCLUSION TEST, OCCLUSION TEST, PRIME TEST AND EXCESSIVE NO DELIVERY ALARM TEST. NO UNEXPECTED EXCESSIVE NO DELIVERY ALARM WAS NOTED. THE INSULIN PUMP WAS RECEIVED WITH A CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, MINOR SCRATCHES ON THE DISPLAY WINDOW, A SCRATCHED RESERVOIR TUBE WINDOW AND A STAINED END CAP STICKER.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER HAD HIGH BLOOD GLUCOSE LEVELS OF 560 MG/DL. THE CUSTOMER REPORTED RECEIVING MULTIPLE NO DELIVERY ALARMS FROM THE INSULIN PUMP. IT WAS REPORTED THAT THE NO DELIVERY ALARMS FROM THE INSULIN PUMP WERE RESOLVED BY A COMPLETE INFUSION SET CHANGE. THE CUSTOMER FURTHER REPORTED THAT THE INFUSION SETS WOULD LAST ONLY ONE AND A HALF DAYS BEFORE GETTING A NO DELIVERY ALARM AGAIN FROM THE INSULIN PUMP. THE CUSTOMER WOULD LIKE TO HAVE THE INSULIN PUMP REPLACED. THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP WOULD BE REPLACED. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO A BACK-UP PLAN PER THE HEALTH CARE PROVIDER'S INSTRUCTION. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
643559 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAS

Patients

Seq Age Sex Outcome Treatment
1 63 YR