FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE S

MDR report key: 4161846 · Received October 10, 2014

Report

Report Number
2649622-2014-11384
Event Type
Injury
Date Received
October 10, 2014
Date of Event
August 15, 2014
Report Date
August 15, 2014
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Removal / Correction Number
Z-0474-2011
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION, BUT RETURNED PRODUCT TESTING FOUND THE DEVICE DID NOT PERFORM AS DESCRIBED IN THE FIELD ACTION.  PRODUCT EVENT SUMMARY: THE DISTAL PORTION OF THE LEAD WAS RETURNED. ANALYSIS WAS PERFORMED AND NO ANOMALIES WERE FOUND. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED APPARENT EXPLANT DAMAGE. CONCOMITANT PRODUCTS: PRODUCT ID 5076-45, LEAD, IMPLANTED: (B)(6) 2005; PRODUCT ID 419388, LEAD, IMPLANTED: (B)(6) 2005. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) AND LEFT VENTRICULAR (LV) LEADS DISLODGED AND HAD HIGH THRESHOLDS. THE LEADS WERE EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
639569 SPRINT QUATTRO SECURE S DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 693558

Patients

Seq Age Sex Outcome Treatment
1 00070 YR Hospitalization| R 5076-45 LEAD, DTBA1D1 ICD