MAXIMO II DR

MDR report key: 4161835 · Received October 10, 2014

Report

Report Number
3004209178-2014-19005
Event Type
Injury
Date Received
October 10, 2014
Date of Event
July 23, 2014
Report Date
April 22, 2015
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
LWS
PMA / PMN Number
P980016
Removal / Correction Number
Z-0113-2011
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION, BUT RETURNED PRODUCT TESTING FOUND THE DEVICE DID NOT PERFORM AS DESCRIBED IN THE FIELD ACTION. PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. THE DEVICE MET 88% OF EXPECTED LONGEVITY. WITHOUT THE HISTORY OF THE PROGRAMMED SETTINGS THROUGHOUT ITS SERVICE LIFE, THERE IS NO WAY TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION. BASED ON THE INFORMATION RECEIVED AND WITHOUT THE RETURN OF THE PRODUCT, IT COULD NOT DETERMINE THIS DEVICE PERFORMED AS DESCRIBED IN THE FIELD ACTION. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE LONGEVITY WAS NOT MEETING THE PATIENT OR CLINICIANS EXPECTATIONS. THE PATIENT WAS NOTED TO BE IN CHRONIC ATRIAL FIBRILLATION AND THE DEVICE IS SENSING ATRIAL FIBRILLATION 100% OF THE TIME. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION, BUT RETURNED PRODUCT TESTING FOUND THE DEVICE DID NOT PERFORM AS DESCRIBED IN THE FIELD ACTION.

Description of Event or Problem · 1

IT WAS ADDITIONALLY REPORTED THAT THE DEVICE REACHED ELECTIVE REPLACEMENT INDICATOR (ERI). THE DEVICE WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
639711 MAXIMO II DR DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MEDTRONIC MED REL MEDTRONIC PUERTO RICO D284DRG

Patients

Seq Age Sex Outcome Treatment
1 00069 YR Hospitalization| R 6947-65 LEAD