SPRINT
Report
- Report Number
- 2649622-2014-11410
- Event Type
- Malfunction
- Date Received
- October 10, 2014
- Date of Event
- August 5, 2014
- Report Date
- August 5, 2014
- Manufacturer
- MPRI
- Product Code
- LWS
- PMA / PMN Number
- P920015
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT EVENT SUMMARY: THE LEAD WAS NOT RETURNED FOR ANALYSIS; HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY INDICATED OVERSENSING ASSOCIATED WITH THE RIGHT VENTRICULAR LEAD THE ANALYST NOTED THE FOLLOWING: BEGINNING (B)(6) 2014, SENSING INTEGRITY COUNTS UP TO 1790; NST EPISODES COLLECTED WITH R-WAVE DOUBLE COUNTING AS SHORT INTERVALS; RV LEAD INTEGRITY WARNING OCCURRED (B)(6) 2014 FOR SENSING INTEGRITY COUNTER GREATER THAN 30 IN 3 DAYS AND VT-NS EPISODES.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT A LEAD INTEGRITY ALERT (LIA) WAS TRIGGERED DUE TO DOUBLE COUNTING OF VENTRICULAR TACHYCARDIA (VT) BY THE RIGHT VENTRICULAR (RV) LEAD DURING SOME NON-SUSTAINED TACHYCARDIA EPISODES. IT WAS FURTHER REPORTED THAT THE RV LEAD DEMONSTRATED CONSISTENTLY HIGH SHORT INTERVAL COUNTS (SIC). THE PHYSICIAN ELECTED NOT TO ADJUST ANY PROGRAMMING AT THIS TIME, AND NO INTERVENTION WAS DONE. THE LEAD REMAINS IN USE AND NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 639781 | SPRINT | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | MPRI | 6932-65 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00063 YR | D314VRG ICD |