FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS

MDR report key: 4161778 · Received October 10, 2014

Report

Report Number
2649622-2014-11425
Event Type
Malfunction
Date Received
October 10, 2014
Date of Event
August 6, 2014
Report Date
August 6, 2014
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT: 559445 LEAD, IMPLANTED: (B)(6) 2010. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WERE CHRONICALLY HIGH RV (RIGHT VENTRICULAR) THRESHOLDS. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
640740 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5076-52

Patients

Seq Age Sex Outcome Treatment
1 00076 YR SEDR01 IPG