FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 4161730 · Received October 10, 2014

Report

Report Number
2649622-2014-11439
Event Type
Injury
Date Received
October 10, 2014
Date of Event
August 14, 2014
Report Date
August 14, 2014
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Removal / Correction Number
Z-0475-2011
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. THERE WAS BLOOD ON THE DISTAL CONDUCTOR OF THE LEAD AND IT WAS NOT OBSTRUCTED. THE ANALYST NOTED THAT AN INSULATION BREACH WAS NOT OBSERVED DURING ANALYSIS. A TRACE AMOUNT OF BLOOD WAS OBSERVED ON THE DISTAL CONDUCTOR. NO OTHER ANOMALIES WERE OBSERVED ON THE RETURNED LEAD PROXIMAL SEGMENT. ALL ELECTRICAL TESTING WAS WITHIN SPECIFIED PARAMETERS. THE FULL LEAD WAS NOT RETURNED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED LOW PACING IMPEDANCE AND LOW HIGH-VOLTAGE IMPEDANCE. IN ADDITION, IT WAS NOTED THAT THERE WAS FLUID IN THE POCKET AND DISCOLORATION UNDER THE LEAD INSULATION INDICATIVE OF COMPROMISED LEAD INSULATION. THE RV LEAD WAS PARTIALLY EXPLANTED, CAPPED AND REPLACED SUCCESSFULLY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
639120 SPRINT QUATTRO SECURE DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 694765

Patients

Seq Age Sex Outcome Treatment
1 00072 YR Hospitalization| R C174AWK ICD