FDA Adverse Event Injury Summary report: N

ATTAIN ABILITY

MDR report key: 4161631 · Received October 10, 2014

Report

Report Number
2649622-2014-11443
Event Type
Injury
Date Received
October 10, 2014
Date of Event
August 1, 2014
Report Date
August 15, 2014
Manufacturer
MPRI
Product Code
OJX
PMA / PMN Number
P080006
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED STRETCHING AND APPARENT EXPLANT DAMAGE.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD EXHIBITED AN INSULATION DEFECT WHICH RESULTED IN DIAPHRAGMATIC/PHRENIC STIMULATION. IT WAS NOTED THAT THE LEAD APPEARED TO BE CUT. THE LEAD WAS PARTIALLY EXPLANTED AND THE DISTAL END REMAINS IN THE PATIENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
640363 ATTAIN ABILITY DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE OJX MPRI 4196

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R