FDA Adverse Event
Injury
Summary report: N
INSYNC III MARQUIS
MDR report key: 4161437
·
Received October 10, 2014
Report
- Report Number
- 9614453-2014-02285
- Event Type
- Injury
- Date Received
- October 10, 2014
- Date of Event
- June 20, 2014
- Report Date
- July 2, 2014
- Manufacturer
- IPG MFG SWITZERLAND
- Product Code
- NIK
- PMA / PMN Number
- P010031
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS EXPLANTED AND REPLACED DUE TO ELECTIVE REPLACEMENT INDICATOR (ERI) WITH UNEXPECTED LONGEVITY DURATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 641502 | INSYNC III MARQUIS | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO | NIK | IPG MFG SWITZERLAND | 7279 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00082 YR | Hospitalization| R |