FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 4161402 · Received October 10, 2014

Report

Report Number
2649622-2014-11532
Event Type
Injury
Date Received
October 10, 2014
Date of Event
July 1, 2014
Report Date
July 1, 2014
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD AN APPARENT FRACTURE WITH HIGH UNIPOLAR AND BIPOLAR IMPEDANCE, AND HIGH UNIPOLAR AND BIPOLAR THRESHOLDS. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
640769 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5076-58

Patients

Seq Age Sex Outcome Treatment
1 00076 YR Hospitalization| R 5076-45 LEAD, ADDR01 IPG