ENRHYTHM MRI
Report
- Report Number
- 9614453-2014-02294
- Event Type
- Injury
- Date Received
- October 10, 2014
- Date of Event
- May 25, 2014
- Report Date
- July 3, 2014
- Manufacturer
- IPG MFG SWITZERLAND
- Product Code
- NVZ
- PMA / PMN Number
- P980035
- Removal / Correction Number
- Z-1440-2010
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GR
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY FOUND THE ERI (ELECTIVE REPLACEMENT INDICATOR) IS THE RESULT OF HIGH INTERNAL BATTERY RESISTANCE. DEVICE ANALYSIS REVEALED THE RETURNED DEVICE INDICATED HIGH IMPEDANCE IN THE BATTERY. THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES; HOWEVER, IT IS SAME/SIMILAR TO A DEVICE MARKETED IN THE U.S. THIS EVENT OCCURRED OUTSIDE THE US AND PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION, BUT RETURNED PRODUCT TESTING FOUND THE DEVICE DID PERFORM AS DESCRIBED IN THE FIELD ACTION. (B)(4).
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) WAS EXPLANTED AND REPLACED DUE TO ELECTIVE REPLACEMENT INDICATOR (ERI), EARLY BATTERY DEPLETION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 641164 | ENRHYTHM MRI | PULSE GENERATOR, PERMANENT, IMPLANTABLE | NVZ | IPG MFG SWITZERLAND | EMDR01 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00084 YR | Hospitalization| R |