FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE S

MDR report key: 4161396 · Received October 10, 2014

Report

Report Number
2649622-2014-11549
Event Type
Malfunction
Date Received
October 10, 2014
Date of Event
July 3, 2014
Report Date
July 3, 2014
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE U.S WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED AND THE HELIX OF THE LEAD WAS EXTRINSICALLY BENT. THE HELIX OF THE LEAD BECAME EXTRINSICALLY DISTORTED DUE TO PULLING/STRETCHING/OVERSTRESS. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT. THE ANALYST NOTED THAT THE LEAD WAS RETURNED WITH THE HELIX BENT AND STRETCHED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN ATTEMPTED IMPLANT, THE RIGHT VENTRICULAR (RV) LEAD BECAME STUCK IN THE TRICUSPID VALVE. DURING AN ATTEMPT TO REPOSITION THE RV LEAD, IT WAS NOTED THAT THE HELIX WAS EXTENDED OUT OF THE TIP OF THE LEAD AND DISENGAGED FROM THE BODY OF THE RV LEAD. THE RV LEAD WAS REMOVED AND REPLACED SUCCESSFULLY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
640767 SPRINT QUATTRO SECURE S DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 6935M62

Patients

Seq Age Sex Outcome Treatment
1