FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 4161335 · Received July 23, 2014

Report

Report Number
3008642652-2014-02303
Event Type
Malfunction
Date Received
July 23, 2014
Date of Event
June 25, 2014
Report Date
July 21, 2014
Manufacturer
ZOLL MANUFACTURING CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. UPON EVALUATION, THE CABLE CONNECTED THE REAR THERAPY ELECTRODE (TE) AND THE DISTRIBUTION NODE (DN) WAS PULLED FROM THE DN STRAIN RELIEF, DAMAGING INTERNAL WIRES. THE ROOT CAUSE OF THE DAMAGED CABLE AND WIRES CANNOT BE POSITIVELY IDENTIFIED, BUT IS LIKELY EXCESSIVE FORCE PLACED ON THE CABLE. NO ADVERSE EVENT RESULTED FROM THE DAMAGED CABLE AND WIRES. THE LAST PATIENT TO USE THIS BELT DID NOT REPORT ANY DEFICIENCIES.

Description of Event or Problem · 1

REVIEW OF SERVICE DATA DETECTED A REPORTABLE PROBLEM. DURING SERVICING, BELT SN (B)(4) FAILED INCOMING HI-POT TESTING. THE LAST PATIENT TO USE THIS BELT DID NOT REPORT ANY DEFICIENCIES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
430721 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL MANUFACTURING CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 NA