FDA Adverse Event Malfunction Summary report: N

CARELINK

MDR report key: 4161322 · Received October 10, 2014

Report

Report Number
2182208-2014-02709
Event Type
Malfunction
Date Received
October 10, 2014
Date of Event
July 14, 2014
Report Date
July 14, 2014
Manufacturer
MEDTRONIC, INC.
Product Code
KRG
PMA / PMN Number
P890003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: ANALYSIS CONFIRMED ERROR AT BOOT UP; PROGRAMMER BOOTS TO A SYSTEM ERROR. ANALYSIS ALSO FOUND THE KEYBOARD SLIDE COVER IS MISSING. LEFT KEYBOARD HINGE AND KEYBOARD LATCH ARE BROKEN. STYLUS TIP IS PULLING AWAY FROM MAIN PEN BODY. CONCOMITANT PRODUCT: 229047 ANALYZER. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PROGRAMMER RECEIVED AN ERROR DURING BOOT UP. THE PROGRAMMER WAS RETURNED FOR REPAIR. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
638160 CARELINK PROGRAMMER, PACEMAKER KRG MEDTRONIC, INC. 2090W

Patients

Seq Age Sex Outcome Treatment
1 2067L RADIO FREQUENCY HEAD