FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE S

MDR report key: 4161320 · Received October 10, 2014

Report

Report Number
2649622-2014-11566
Event Type
Injury
Date Received
October 10, 2014
Date of Event
July 16, 2014
Report Date
July 16, 2014
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Removal / Correction Number
Z-0474-2011
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY INDICATED THE IMPEDANCE ON THE RV (RIGHT VENTRICULAR) DEFIBRILLATION COIL WAS BEYOND THE EXPECTED UPPER RANGE AND A LEAD IMPEDANCE OUT OF RANGE ALERT. AN OUT OF TOLERANCE SUBTHRESHOLD LEAD IMPEDANCE ALERT OCCURRED ON (B)(6) 2014. MAXIMUM DEFIBRILLATION ACTIVE CAN IMPEDANCE RISES FROM 63 OHM ON (B)(6) 2014 TO GREATER THAN 250 OHM ON (B)(6) 2014.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE A SIMILAR MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT AFTER THE ROUTINE CHANGE OUT OF THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) THE VENTRICULAR LEAD IMPEDANCE WAS UNSTABLE AND ARTIFACT WAS NOTED. LEAD FRACTURE WAS SUSPECTED. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
637720 SPRINT QUATTRO SECURE S DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 693565

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R