FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 4161122 · Received October 10, 2014

Report

Report Number
2649622-2014-11627
Event Type
Malfunction
Date Received
October 10, 2014
Date of Event
November 1, 2013
Report Date
July 15, 2014
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE ACTUAL PRODUCT WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA WAS COLLECTED FROM THE DEVICE AND WAS ANALYZED. ANALYSIS OF THE DEVICE MEMORY INDICATED THE IMPEDANCE ON THE RV (RIGHT VENTRICULAR) DEFIBRILLATION COIL WAS BEYOND THE EXPECTED UPPER RANGE. ANALYSIS OF THE DEVICE MEMORY INDICATED THE IMPEDANCE ON THE SVC (SUPERIOR VENA CAVA) DEFIBRILLATION COIL WAS BEYOND THE EXPECTED UPPER RANGE. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SUPERIOR VENA CAVA (SVC) AND RIGHT VENTRICULAR (RV) DEFIBRILLATION IMPEDANCE HAS BEEN HIGH. THERE HAVE BEEN NO ALERTS BECAUSE ALL OF THE ALERTS WERE PROGRAMMED OFF. A FRACTURE IS SUSPECTED. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS FURTHER REPORTED THAT THE RV LEAD CONTINUED TO EXHIBIT HIGH IMPEDANCE. THE LEAD WAS CAPPED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
637888 SPRINT QUATTRO SECURE DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 694765

Patients

Seq Age Sex Outcome Treatment
1 00063 YR Hospitalization| R 7232CX IPG