FDA Adverse Event Injury Summary report: N

ENRHYTHM MRI

MDR report key: 4161085 · Received October 10, 2014

Report

Report Number
9614453-2014-02320
Event Type
Injury
Date Received
October 10, 2014
Date of Event
April 18, 2014
Report Date
July 1, 2014
Manufacturer
IPG MFG SWITZERLAND
Product Code
NVZ
PMA / PMN Number
P980035
Removal / Correction Number
Z-1440-2010
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY INDICATED THE BATTERY IMPEDANCE VALUE WAS BEYOND THE EXPECTED UPPER RANGE. THE ANALYST COMMENTED THAT ELECTIVE REPLACEMENT INDICATOR (ERI) DUE TO HIGH BATTERY IMPEDANCE WAS RECORDED ON (B)(6) 2014, PRIOR TO EXPLANT. RAMWARE VERSION 3 WAS INSTALLED ON (B)(6) 2011. THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION. BASED ON THE INFORMATION RECEIVED AND WITHOUT THE RETURN OF THE PRODUCT, IT COULD NOT DETERMINE THIS DEVICE PERFORMED AS DESCRIBED IN THE FIELD ACTION.

Additional Manufacturer Narrative · 1

THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES, HOWEVER, IT IS SIMILAR TO A DEVICE MARKETED IN THE U.S. THE EVENT IS BEING REPORTED DUE TO AN ALLEGED MALFUNCTION. (B)(4). THIS DEVICE WAS INCLUDED IN THE FIELD ACTION. BASED ON THE INFORMATION RECEIVED AND WITHOUT THE RETURN OF THE PRODUCT, IT COULD NOT BE DETERMINED IF THIS DEVICE PERFORMED AS DESCRIBED IN THE FIELD ACTION.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE REACHED RECOMMENDED REPLACEMENT TIME (RRT) AFTER FOUR YEARS AND A DEVICE EXPLANT IS PLANNED. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
641356 ENRHYTHM MRI PULSE GENERATOR, PERMANENT, IMPLANTABLE NVZ IPG MFG SWITZERLAND EMDR01

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R